MECG 2.0心電數(shù)據(jù)庫播放器
全模擬訊號輸出的ECG數(shù)據(jù)庫播放器,協(xié)助您完成系統(tǒng)及算法驗證,適用于研發(fā)及法規(guī)符合性測試??芍苯蛹虞d及撥放各類型心電數(shù)據(jù)庫,符合國際及國內(nèi)最新版標準:IEC 60601-2-25、IEC 60601-2-47、ANSI/AAMI EC57、YY0782、YY0885
產(chǎn)品簡介
全模擬訊號輸出的ECG數(shù)據(jù)庫播放器,協(xié)助您完成系統(tǒng)及算法驗證,適用于研發(fā)及法規(guī)符合性測試??芍苯蛹虞d及撥放各類型心電數(shù)據(jù)庫,符合國際及國內(nèi)新版標準:IEC 60601-2-25、IEC 60601-2-47、ANSI/AAMI EC57、YY0782、YY0885
產(chǎn)品特點
-
可供心電診斷、可攜式心電設備仿真數(shù)據(jù)庫產(chǎn)生器
-
支持12 導聯(lián)心電圖機,8 個信道同步產(chǎn)生數(shù)據(jù)庫訊號
-
內(nèi)建CAL、ANE 及Biological 波形
-
可直接加載及播放AHA、MIT、CU 數(shù)據(jù)庫,無須格式轉換
-
較佳的輸出訊號準確度,電壓輸出<±8μV
-
提供軟件開發(fā)工具包(Software Development Kit) 讓用戶自行開發(fā)全自動化測試軟件
產(chǎn)品規(guī)格
Item | Details / Reference | Value |
---|---|---|
Output channels |
The 8 output channels are provided through a network as specified in IEC 60601-2-51 to provide signals to 10 lead electrodes; in the device under test, this will be displayed as 12 leads. |
8 outputs 10 lead electrodes 12 leads |
Voltage accuracy |
IEC 60601-2-51 specifies a limit of ±1%, but does not provide a lower limit (all systems must have a lower limit). An inferred specification of 1% ±5μV is derived from the device under test specification in IEC 60601-2-51 of 5% ±25μV. |
±1% for |
Output voltage |
The output voltage on most of the database and ECG is +5mV~-5mV. |
±5mV |
Output noise level 0-150Hz |
Output noise should not influence the test. A value a 5μV is suitable for this requirement. Can be verified by monitoring the signal in the device under test using a “diagnostic” filter setting. |
<5μV |
Time accuracy |
IEC 60601-2-51 does not provide any limits. An inferred limit from the device under test. An inferred limit of ±1% is used (see 4.2). The system’s design accuracy exceeds 0.1% as a 100ppm crystal reference is used. |
±1% |
Sampling rate |
A maximum sampling rate of 1kHz matches the sampling rates of ECG files. |
1kHz (8 channels) |
Power supply | Powered from the USB supply (5V 0.2A) | N/A |
Environment |
Intended for normal laboratory environment. The selection of critical components such as reference voltages, DAC, precision resistors are known to be stable in the range shown. |
15-30°C 10-95% RH |